Target Mineral Quant (TMQ)
One of the most powerful features of AMICS is its ability to link x-ray chemical analysis to a spatial location on any specimen. The Target Mineral Quant (TMQ) function leverages this to perform corrected quantification at the particle or grain level, providing detailed understanding of elemental deportment.
Key capabilities:
- Filter and sort grains to create summed spectra per grain
- Export grain tables with coordinates and elemental composition
- Align with fiducial points for integration with third-party tools (LA-ICP-MS, SIMS)
Accessing TMQ
Open TMQ via the TMQ tab above the main data visualization panel. The TMQ ribbon context tab appears with dedicated controls.
{{FIG: TMQ tab showing filtered grains in upper panel, rejected grains in lower panel, and filter parameters on the right | TMQ tab overview}}
Filter Parameters
Filter grains by the following criteria:
Filter | Description |
|---|---|
Having Material | Isolate grains containing one or more specified materials |
Minimum Area | Grains with total area greater than this value (µm²) |
Maximum Area | Grains with total area less than this value (µm²) |
Minimum BSE | Grains with average BSE greater than this value |
Maximum BSE | Grains with average BSE less than this value |
Min Radius (µm) | Grains with radius greater than this value (farthest distance from centroid) |
Filter by Count | Enable filtering by minimum grain count of target material per particle |
Mineral Count | Minimum number of target material grains required in a particle |
Click Filter! to apply. Grains passing the filter appear in the upper panel; rejected grains appear in the lower panel.
Chemistry Types (New in 5.1)
Select the chemistry display mode from the TMQ ribbon → Quantification panel → Type dropdown:
Type | Description |
|---|---|
Weight % | Elemental weight percent (default) |
Atomic % | Elemental atomic percent |
Weight Oxide | Oxide weight percent using oxide formulas from the Materials Library |
Quantification
- Load quantification methods: References tab → Quantification → Update Method List
- Select the recipe from the EDS Methods dropdown on the TMQ ribbon
- Click Quant Selected (selected grains only) or Quant All (all filtered grains)
- Results appear as additional element columns in the Quantification Result panel
Normalization
Enable the Normalize checkbox to normalize elemental concentrations to 100%.
Decimal Precision
Adjust decimal places using the Decimal Point field on the TMQ ribbon.
Quantification Result Panel
Column | Description |
|---|---|
Center Position | Cartesian coordinate of the calculated centroid based on x-ray point positions |
Center Reference | Alternative coordinate for fiducial-point realignment with third-party tools |
Aspect Ratio | Shape factor — ratio of minimum to maximum orthogonal diameter |
Area (Pixels) | Number of pixels within the grain/particle |
BSE (Max) | Highest BSE value measured across all pixels |
Xrays | Number of x-ray points used to calculate the summed spectrum |
Total Count | Total count of the summed spectrum |
Element columns | Quantified element concentrations (after Quant is run) |
Reticle Controls
The reticle marks the target collection location on each grain. Control from the TMQ ribbon → Reticle panel:
Button | Action |
|---|---|
Move Up / Down / Left / Right | Nudge the reticle position |
The reticle auto-fits to the position covering the maximum number of x-ray points.
Undo and Redo (New in 5.1)
Undo/redo for reject operations: TMQ ribbon → Reject panel → Undo / Redo buttons.
Keyboard Navigation
Action | Description |
|---|---|
Previous / Next | Navigate between grains |
Focus | Center view on selected grain |
Toggle Mark | Mark or unmark current grain |
Pick / Unpick | Select or deselect current grain |
Inverse Selection | Invert selection |
Deselect All | Clear all selections |
Ctrl+E | Export data to CSV/TSV |
Export Options
Export from the TMQ ribbon → Export Options panel:
Button | Description |
|---|---|
Selected Spectra | Export SPX spectra for selected grains |
All Spectra | Export SPX spectra for all filtered grains |
Selected Images | Export particle images for selected grains |
All Images | Export particle images for all filtered grains |
Data Export | Export results table as CSV/TSV (Ctrl+E) |
Object Name | Include grain names in export |
Data can also be copied to clipboard using the copy button on the Quantification Result panel.
Sending Particles to TMQ
From the Feature Grid, right-click selected particles → Send Selected Particles to TMQ.
Saving and Restoring Filter Parameters
- Save Search! — Save current filter parameters with a name
- Restore Search! — Load previously saved parameters
Parameters are stored as JSON. When transferring between measurements, ensure both use the same Classification Workflow, as material IDs are workflow-specific.
Integration with LA-ICP-MS
TMQ integrates with third-party analytical tools via fiducial points:
- Mark fiducial points on the specimen in AMICS Investigator
- Run measurement and apply TMQ filtering
- Export grain coordinates with fiducial reference points
- Import coordinates into LA-ICP-MS or SIMS software for targeted analysis
See [Fiducial Points] for setup instructions.
See Also
- [Feature Grid Tab] — Selecting particles for TMQ analysis
- [Elemental Visualization] — Complementary element distribution view
- [Fiducial Points] — Setting up reference points for third-party tools